4.3 Article

Impact of monitoring blood volume in the BD BACTEC™ FX blood culture system: virtual volume versus actual volume

期刊

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.diagmicrobio.2014.11.001

关键词

Blood culture; Blood volume; Monitoring; Aerobic bottle; Anaerobic bottle

向作者/读者索取更多资源

The BACTEC (TM) FX (BD, USA) has been recently upgraded to measure virtual blood volume (W) in batches of 25 PLUS Aerobic/F bottles. VVs were compared with manually measured actual volumes (AVs). The mean AV was larger in anaerobic bottles: 2.6 +/- 1.9 mL in 2995 Aerobic/F bottles and 3.3 +/- 1.8 mL in 3053 Lytic/10 Anaerobic/F bottles (P< 0.001). The VV of Aerobic/F bottles was 2.4 +/- 1.4 mL, and there was a significant correlation between the Ws and AVs of each department (P < 0.001). The potential positivity in aerobic bottles increased 1.2-fold with a 1-mL increase in the AV (P < 0.001). AV and VV were consistently less than adequate. Although the proportional effect of volume on positivity can be proven only with the AV, VVs are a reliable substitute for AVs to monitor the overall blood volumes and provide feedback to the individual departments or locations. (C) 2015 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据