4.6 Article

Improved methods for nanoindentation Berkovich probe calibrations using fused silica

期刊

JOURNAL OF MATERIALS SCIENCE
卷 53, 期 7, 页码 4814-4827

出版社

SPRINGER
DOI: 10.1007/s10853-017-1922-8

关键词

-

向作者/读者索取更多资源

Improved methods are needed to improve the reliability and reporting of Berkovich probe calibrations. Current methods for calibrating machine compliance rely substantially on the subjectivity of the user to choose and exclude from the analysis nanoindentations below a threshold size that are affected by probe tip imperfections or the fused silica surface properties. Furthermore, it can be difficult to identify erroneous experimental data, such as those resulting from dirty or damaged probes. To help overcome these issues in Berkovich probe calibrations, a pre-nanoindentation liftoff analysis and systematic analysis of the Stone-Yoder-Sproul plot were developed. The utility of the analyses was demonstrated through analysis of experimental data. After correcting the fused silica data for machine compliance, the probe area function calibration was performed following established protocols. Finally, a concise protocol for reporting Berkovich probe calibration results was established to improve comparisons between experiments performed with different Berkovich probes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据