4.5 Article

Adaptable setups for magnetic drug targeting in human muscular arteries: Design and implementation

期刊

JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS
卷 438, 期 -, 页码 173-180

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jmmm.2017.04.058

关键词

Magnetic targeting; Drug delivery; Muscular vessel; Particle tracing; In-vitro validation

向作者/读者索取更多资源

Magnetic drug targeting has been used to steer magnetic therapeutic agents and has received much attention for capillaries and human brain arteries. In this paper, we focus on noninvasive targeting of nanoparticles in muscular arteries, in where the vessel diameter and blood flow are much challengingly higher than brain capillaries. We aim to design a low intensity magnetic field which avoids potential side effects on blood cells while steers particles with high targeting rate. The setup design procedure is considerably flexible to be used in a wide variety of large vessels. Using particle tracing, a new method is proposed to connect the geometry of the vessel under the action of targeting to the required magnetic force. Specifications of the coil which is placed outside the body are derived based on this required force. Mutual effects of coil dimensions on the produced magnetic force are elaborated and summarized in a design flowchart to be used for arbitrary muscular vessel sizes. The performance of the optimized coil is validated by in vitro experiments and it is shown that particles are steered with the average efficiency of 80.2% for various conditions. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据