Comparison of fast acquisition strategies in whole-heart four-dimensional flow cardiac MR: Two-center, 1.5 Tesla, phantom and in vivo validation study

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 47, 期 1, 页码 272-281

出版社

WILEY
DOI: 10.1002/jmri.25746

关键词

4D flow cardiac MR; phase-contrast magnetic resonance imaging; MR flow imaging; flow quantification; validation

资金

  1. British Heart Foundation [FS/10/62/28409]
  2. Dutch ZonMw [104003001]

向作者/读者索取更多资源

PurposeTo validate three widely-used acceleration methods in four-dimensional (4D) flow cardiac MR; segmented 4D-spoiled-gradient-echo (4D-SPGR), 4D-echo-planar-imaging (4D-EPI), and 4D-k-t Broad-use Linear Acquisition Speed-up Technique (4D-k-t BLAST). Materials and MethodsAcceleration methods were investigated in static/pulsatile phantoms and 25 volunteers on 1.5 Tesla MR systems. In phantoms, flow was quantified by 2D phase-contrast (PC), the three 4D flow methods and the time-beaker flow measurements. The later was used as the reference method. Peak velocity and flow assessment was done by means of all sequences. For peak velocity assessment 2D PC was used as the reference method. For flow assessment, consistency between mitral inflow and aortic outflow was investigated for all pulse-sequences. Visual grading of image quality/artifacts was performed on a four-point-scale (0=no artifacts; 3=nonevaluable). ResultsFor the pulsatile phantom experiments, the mean error for 2D PC=1.01.1%, 4D-SPGR=4.91.3%, 4D-EPI=7.61.3% and 4D-k-t BLAST=4.41.9%. In vivo, acquisition time was shortest for 4D-EPI (4D-EPI=8 +/- 2min versus 4D-SPGR=9 +/- 3min, P<0.05 and 4D-k-t BLAST=9 +/- 3min, P=0.29). 4D-EPI and 4D-k-t BLAST had minimal artifacts, while for 4D-SPGR, 40% of aortic valve/mitral valve (AV/MV) assessments scored 3 (nonevaluable). Peak velocity assessment using 4D-EPI demonstrated best correlation to 2D PC (AV:r=0.78, P<0.001; MV:r=0.71, P<0.001). Coefficient of variability (CV) for net forward flow (NFF) volume was least for 4D-EPI (7%) (2D PC:11%, 4D-SPGR: 29%, 4D-k-t BLAST: 30%, respectively). ConclusionIn phantom, all 4D flow techniques demonstrated mean error of less than 8%. 4D-EPI demonstrated the least susceptibility to artifacts, good image quality, modest agreement with the current reference standard for peak intra-cardiac velocities and the highest consistency of intra-cardiac flow quantifications. Level of Evidence: 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2018;47:272-281.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据