3.8 Article

Determination of ethambutol in biological samples using graphene oxide based dispersive solid-phase microextraction followed by ion mobility spectrometry

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s12127-019-00253-z

关键词

Ethambutol; Microextraction; Graphene oxide; Ion mobility spectrometry; Biological samples

资金

  1. Damghan University Research Council

向作者/读者索取更多资源

A simple, fast and reliable graphene oxide nanosheets based dispersive solid-phase microextraction methodology was described for the quantification of trace amount of ethambutol. The determination of ethambutol quantified with the aim of ion mobility spectroscopy as a sensitive, rapid, inexpensive and environmentally friendly instrument. Effects of relevant experimental parameters on the method efficiency such as pH, type of buffer and its volume, amount of absorbent, desorption solvent and extract time were investigated to reach the maximum efficiency of the proposed method. Under the optimum conditions, the calibration curve was linear in the range of 1 to 120 mu g L-1 with the R-squared (R-2) of 0.9990. The limit of detection for proposed method (n = 8) was 0.4 mu g L-1 and the relative standard deviations were obtained (n = 8) 3.3% and 1.6% for 10 and 100 mu g L-1, respectively. The proposed method was successfully applied for the preconcentration and determination of ethambutol in the different biological samples such as plasma, saliva, breast milk and artificial tear.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据