Performance of the Dimension TAC assay and comparison of multiple platforms for the measurement of tacrolimus

期刊

出版社

WILEY
DOI: 10.1002/jcla.22357

关键词

Dimension TAC assay; immunosuppressive agent; performance evaluation; tacrolimus; transplant

向作者/读者索取更多资源

Background: Therapeutic monitoring of tacrolimus is essential for reducing organ rejection and adverse effects. The measurement of tacrolimus in whole blood is taken by many automated platforms. We evaluated the analytical performance of the Dimension TAC assay, which is an upgraded reagent from the previous Dimension TACR assay. Methods: The evaluations involved determination of precision, linearity, detection capability, and reagent lot-to-lot variability between three lot numbers. Correlation studies were conducted using the Dimension TACR assay, Architect, Elecsys assay, and MassTrak LC-MS/MS. Results: The total coefficient of variation was below 10%. Acceptable linearity was observed in their respective reportable ranges. The limit of blank, limit of detection, and limit of quantification were 0.29, 0.47, and 0.81 ng/mL, respectively. Correlation analysis indicated that the Dimension TAC assay results were comparable to that of the Dimension TACR assay, Architect, and Elecsys results in liver and heart transplant patients. In kidney transplant patients, the Dimension TAC assay showed the poor correlation with Architect and Elecsys. The results from these assays were slightly higher than that of MassTrak. We found little lot-to-lot reagent variation among the reagents evaluated. Conclusion: The overall analytical performance of the Dimension TAC assay is acceptable for therapeutic monitoring in clinical practice. Our study that compared different platforms may provide some useful information regarding which test method to use.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据