Evaluation of intensity drift correction strategies using MetaboDrift, a normalization tool for multi-batch metabolomics data

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1523, 期 -, 页码 265-274

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2017.09.023

关键词

Metabolomics; LC-MS; Intensity drift; Drift correction; Normalization; Meal tolerance test

资金

  1. National Institutes of Health [DK092558, DK099034]
  2. NIH [DK089503]
  3. Michigan Regional Comprehensive Metabolomics Resource Core [DK097153]
  4. A. Alfred Taubman Medical Research institute
  5. Robert C. and Veronica Atkins Foundation

向作者/读者索取更多资源

In recent years, mass spectrometry-based metabolomics has increasingly been applied to large-scale epidemiological studies of human subjects. However, the successful use of metabolomics in this context is subject to the challenge of detecting biologically significant effects despite substantial intensity drift that often occurs when data are acquired over a long period or in multiple batches. Numerous computational strategies and software tools have been developed to aid in correcting for intensity drift in metabolomics data, but most of these techniques are implemented using command-line driven software and custom scripts which are not accessible to all end users of metabolomics data. Further, it has not yet become routine practice to assess the quantitative accuracy of drift correction against techniques which enable true absolute quantitation such as isotope dilution mass spectrometry. We developed an Excel-based tool, MetaboDrift, to visually evaluate and correct for intensity drift in a multi-batch liquid chromatography - mass spectrometry (LC-MS) metabolomics dataset. The tool enables drift correction based on either quality control (QC) samples analyzed throughout the batches or using QC-sample independent methods. We applied MetaboDrift to an original set of clinical metabolomics data from a mixed-meal tolerance test (MMTT). The performance of the method was evaluated for multiple classes of meta-bolites by comparison with normalization,using isotope-labeled internal standards. QC sample-based intensity drift correction significantly improved correlation with IS-normalized data, and resulted in detection of additional metabolites with significant physiological response to the MMTT. The relative merits of different QC-sample curve fitting strategies are discussed in the context of batch size and drift pattern complexity. Our drift correction tool offers a practical, simplified approach to drift correction and batch combination in large metabolomics studies. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据