Stir bar sorptive extraction and liquid chromatography-tandem mass spectrometry determination of polar and non-polar emerging and priority pollutants in environmental waters

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1500, 期 -, 页码 43-52

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2017.04.007

关键词

Stir bar sorptive extraction; Liquid chromatography-tandem mass spectrometry; Polar and non-polar emerging and priority pollutants; Ethylene glycol modified silicone stir bar coating; Surface water; Tap water

资金

  1. Spanish Ministry of Economy and Competitiveness [CGL2013-44402-R]

向作者/读者索取更多资源

An analytical method based on stir bar sorptive extraction (SBSE) was developed and validated for the determination of environmental concern pollutants in environmental waters by liquid chromatography-tandem mass spectrometry (LC-MS/MS). Target compounds include six water and oil repellents (perfluorinated compounds), four preservatives (butylated hydroxytoluene and three parabens), two plasticizers (bisphenol A and di(2-ethylhexyl)phthalate), seven surfactants (four linear alkylbenzene sulfonates, nonylphenol and two nonylphenol ethoxylates), a flame retardant (hexabromocyclododecane), four hormones, fourteen pharmaceutical compounds, an UV-filter (2-ethylhexyl 4-methoxycinnamate) and nine pesticides. To achieve the simultaneous extraction of polar and non-polar pollutants two stir bar coatings were tested, the classic polydimethylsiloxane (PDMS) coating and the novel ethylene glycol modified silicone (EG-silicone). The best extraction recoveries were obtained using EG-silicone coating. The effects of sample pH, volume and ionic strength and extraction time on extraction recoveries were evaluated. The analytical method was validated for surface water and tap water samples. The method quantification limits ranged from 7.0 ng L-1 to 177 ng L-1. The inter-day precision, expressed as relative standard deviation, was lower than 20%. Accuracy, expressed as relative recovery values, was in the range from 61 to 130%. The method was applied for the determination of the 48 target compounds in surface and tap water samples. (C) 2017 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据