期刊
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 35, 期 2, 页码 357-365出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c9ja00377k
关键词
-
资金
- National Key Research and Development Program of China [2017YFF0106202]
- Key Research Program of Frontier Sciences, CAS [QYZDJ-SSW-JSC037]
- LiaoNing Revitalization Talents Program [XLYC1807110]
- Youth Innovation Promotion Association, CAS
In the microanalysis of laser-induced breakdown spectroscopy, the influence of surface roughness on spectral stability and quantitative analysis capability was studied for the first time when the laser ablation crater diameter was approximately 10 mu m. By using a microchip laser, the spectra were acquired on a set of carbon steel samples, whose roughness changed from 0.0392 mu m to 0.5248 mu m. Besides, the study found that the surface roughness of the sample neglected the effect on spectral stability via screening and analyzing the spectral data when the sample stage moved at a constant speed. Moreover, when the spectral collection was performed with the ablation crater overlap rate of approximately 75% and Fe line was used as the internal standard, the effect of sample surface roughness on the quantitative analysis of the spectrum was negligible, and the resulting calculated determination coefficients of the Mn, Si, Ni, V, Cr, and Cu elements all exceeded 0.993.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据