期刊
JOURNAL OF ARTHROPLASTY
卷 32, 期 3, 页码 777-782出版社
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2016.08.025
关键词
knee arthroplasty; KneeAlign; accelerometer; conventional guides; computer-assisted surgery
类别
Background: Accelerometer-based, portable navigation devices have been introduced as a less invasive and simpler technique to perform navigated surgical implantation of knee prostheses. They have been postulated to have better accuracy than conventional instruments in restoration of alignment in total knee arthroplasty. Methods: A total of 190 patients were enrolled in this prospective, randomized controlled trial and underwent total knee arthroplasty using either the KneeAlign or conventional guides. Multiplanar alignment was evaluated with a CT imaging protocol. Results: A total of 86.5% of portable navigation device and 82.2% of conventional group had a postoperative hip-knee angle within 3 degrees of neutral alignment (P = .54). There was no significant difference between the 2 groups for component coronal and sagittal plane alignment. Portable navigation device did not significantly increase the time to perform the surgery. Conclusion: Portable navigation device demonstrates accurate restoration of alignment; however, there was no statistically significant difference when compared with conventional guides. (C) 2016 Elsevier Inc. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐