3.8 Article

Assessment of the FilmArray® multiplex PCR system and associated meningitis/encephalitis panel in the diagnostic service of a tertiary hospital

期刊

INFECTION PREVENTION IN PRACTICE
卷 2, 期 2, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.infpip.2020.100042

关键词

Multiplex PCR; Meningitis; encephalitis; Reduced time; Improved patient outcome; Economic benefit

向作者/读者索取更多资源

Rapid and accurate diagnosis of meningitis/encephalitis (M/E) is essential for successful patient outcomes. The FilmArray circle minus meningitis/encephalitis Panel (MEP) is a multiplexed PCR test for simultaneous, rapid detection of pathogens directly from cerebrospinal fluid (CSF) samples. 94 prospectively collected CSF specimens from patients with clinical suspicion of infective M/E underwent testing for 14 pathogens simultaneously, including Escherichia coli, Haemophilus influenzae, Neisseria meningitidis, and Varicella zoster. MEP demonstrated 95% agreement with current PCR methods, resulting in 16 diagnosed cases of M/E. Typically, the FilmArray (R) MEP results were delivered within approximately one hour, contrasting with current practices taking up to 5.6 days. Given the significant morbidity and mortality associated with delayed diagnosis of central nervous system infections, the FilmArray (R) MEP is a useful addition to the diagnostic capabilities of a clinical microbiology department.(C) 2020 The Authors. Published by Elsevier Ltd on behalf of The Healthcare Infection Society. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据