Characterization of a commercially-available, low-pressure UV lamp as a disinfection system for decontamination of common nosocomial pathogens on N95 filtering facepiece respirator (FFR) material

期刊

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/d0ew00404a

关键词

-

资金

  1. NSERC COVID-19 Alliance Grant
  2. NSERC/Halifax Water Industrial Research Chair program

向作者/读者索取更多资源

A commercially-available UV surface disinfection system used for hospital room disinfection was characterized for use as a UV-inactivation system for N95 filtering facepiece respirator (FFR) respirators. The output of light was initially characterized and assessed for its ability to penetrate through the individual N95 respirator layers. Following characterization, disinfection performance was tested using a range of model pathogens. In a series of experiments, coupons of respirator material were inoculated withEscherichia coli,Staphylococcus aureus,Geobacillus stearothermophilusspores,Bacillus cereus,Pseudomonas aeruginosa, and influenza A virus. Inoculated coupons were then treated with UV light from a low-pressure UV disinfection system during specific treatment times. Microbial enumeration was performed pre- and post-UV treatment. It was found that respirator coupons inoculated withE. coli, influenza A,P. aeruginosa,B. cereusandG. stearothermophilusresulted in below detection counts when exposed to UV fluences between 410 to 2700 mJ cm(-2). In addition, UV light was unable to sufficiently penetrate all layers of respirator material, highlighting the importance of flipping respirators when exposing them to UV light. This work suggests that with further validation, UV light will be a promising touchless treatment option for frontline professionals who need to repurpose N95 respirators for subsequent use.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据