Monocular Vision-Based Earth's Graviation Method Used for Low-Frequency Vibration Calibration

期刊

IEEE ACCESS
卷 8, 期 -, 页码 129087-129093

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/ACCESS.2020.3009294

关键词

Acceleration; Sensitivity; Calibration; Accelerometers; Image edge detection; Synchronization; Uncertainty; Low-frequency accelerometer; sensitivity phase and magnitude; monocular vision; time-spatial synchronization (TSS); laser interferometry

资金

  1. National Key Research and Development Program of China [2017YFF0205003]
  2. Quality and Technical Supervision Ability Promotion Project [ANL1820]
  3. National Natural Science Foundation of China [51605461]

向作者/读者索取更多资源

At present, the sensitivity phase of low-frequency accelerometer is commonly calibrated by time synchronization (TS), which needs to strictly align its input excitation acceleration signal and output signal in the time domain. However, TS is very difficult to be implemented and has severely restricted the improvement of the measurement accuracy. A novel calibration method that combines the monocular vision method and time-spatial synchronization technique is investigated to achieve the high-accuracy sensitivity phase calibration. The sensitivity phase is accurately calibrated by determining the aligned spatial position between the excitation acceleration signal and the output signal with the monocular vision method. The sensitivity magnitude can also be simultaneously calibrated. Experimental results show that the calibrated sensitivity phase and magnitude by the investigated method agree well with those by the laser interferometry in the range from 0.3 Hz to 2 Hz. The calibration accuracy of the investigated method is especially superior to that of the laser interferometry in the range from 0.01 Hz to 0.3 Hz.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据