4.4 Article

Spectral DQE of the Volta phase plate

期刊

ULTRAMICROSCOPY
卷 218, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.ultramic.2020.113079

关键词

-

资金

  1. US National Institutes of Health [5 R01 GM126011]
  2. Japan Science and Technology Agency (JST) PRESTO [18069571]

向作者/读者索取更多资源

The Volta Phase Plate (VPP) consists of a heated, thin film that is placed in the same plane as the focused diffraction pattern of an electron microscope. A change in surface potential develops at the point irradiated by the intense, unscattered electron beam, and this altered surface potential produces, in turn, a phase shift between the unscattered and scattered parts of the electron wave. While the VPP thus increases the image contrast for weak-phase objects at low spatial frequencies, we report here that it also leads to the loss of an increasing fraction of the signal at higher resolution. The approximately linear dependence (with increasing resolution) of this loss has been quantified at 200 kV and 300 kV, using evaporated-carbon films of different thicknesses as Volta phase plates. In all cases, the loss of signal remains almost independent of variation of the conditions and parameters that were tested. In spite of having done a number of additional, discovery-based experiments, the cause of this loss of signal remains unexplained at this point.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据