Development of an HDR-BT QA tool for source position verification

期刊

出版社

WILEY
DOI: 10.1002/acm2.13063

关键词

applicator modeling; applicator offset; HDR; QA phantom; source position

资金

  1. JSPS KAKENHI [15K19207]
  2. Grants-in-Aid for Scientific Research [15K19207] Funding Source: KAKEN

向作者/读者索取更多资源

Purpose: This study aimed to develop a high-dose-rate brachytherapy (HDR-BT) quality assurance (QA) tool for verification of source positions, and to report on its effectiveness. Methods We fabricated a cuboid phantom measuring 30 x 30x3 cm(3) with spaces to embed Fletcher-Williamson tandem and ovoid applicators. Lead-based, cylindrically shaped radiopaque markers, which scatter radiation and blacken the Gafchromic (R) RTQA2 films placed on the applicators, were inserted into the phantom to determine the applicator tip and reference source positions. A three-dimensional image-guided brachytherapy (3D-IGBT) plan was generated, and the source positions on the film and radiation treatment planning system (RTPS) were verified with the tool. Source position errors were evaluated as the distance in the applicator axis direction between the source position and the center position of two radiopaque marker pairs. Results Source position errors on the film and RTPS were in good agreement with one another and were all within 0.5 mm for all applicators. Offset values of each applicator were in good agreement with the value determined in treatment planning (6 mm). The expanded measurement uncertainty of our QA tool was estimated to be 0.87 mm, with a coverage factor k of 2. Conclusions Our new HDR-BT QA tool developed for comprehensive source position verification will be useful for cross checking actual source positions and planned source positions on the RTPS.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据