期刊
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL
卷 64, 期 8, 页码 1161-1174出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TUFFC.2017.2706189
关键词
Acoustic; computational acoustic tomography; imaging; inverse scattering; split step Fourier method; 2-D; 3-D
资金
- National Cancer Institute, NIH through the project Partnership 3: multimode quantitative ultrasound prototype breast scanner [R01-CA138536]
We describe a novel 3-D ultrasound technology, the quantitative transmission ultrasound system and algorithm to image a pendent breast in a water bath. Quantitative accuracy is verified using phantoms. Morphological accuracy is verified using cadaveric breast and in vivo images, and spatial resolution is estimated. This paper generalizes an earlier 2-D algorithm to a full 3-D inversion algorithm and shows the importance of such a 3-D algorithm for artifact suppression as compared with the 2-D algorithm. The resultant high-resolution ultrasound images, along with quantitative information regarding tissue speed of sound/stiffness, provide a more accurate depiction of the breast anatomy and lesions, contributing to improved breast care.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐