4.7 Article

A Compressed Sensing Strategy for Synthetic Transmit Aperture Ultrasound Imaging

期刊

IEEE TRANSACTIONS ON MEDICAL IMAGING
卷 36, 期 4, 页码 878-891

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TMI.2016.2644654

关键词

Compressed sensing; contrast; frame rate; resolution; synthetic transmit aperture; ultrasound imaging

资金

  1. National Natural Science Foundation of China [61322101, 61271131, 81471665, 81561168023]
  2. National Key R&D Program of China [2016YFC0102201, 2016YFC0104705]

向作者/读者索取更多资源

A novel beamforming technique, named compressed sensing based synthetic transmit aperture (CS-STA) is proposed to speed up the acquisition of ultrasound imaging. This technique consists of three steps. First, the ultrasound transducer transmits randomly apodized plane waves for a number of times andreceives the backscattered echoes. Second, the recorded backscattered echoes are used to recover the full channel dataset of synthetic transmit aperture (STA) with a compressed sensing (CS) reconstruction algorithm. Finally, an STA image is beamformed from the recovered full STA dataset. As CS allows recovering a signal from its few linear measurements with high probability, CS-STA is capable of recovering the STA image with fewer firings (i.e., higher frame rate) and retaining the high resolution of STA. In addition, the contrast of the STA image can be improved at the same time owing to the higher energy of plane wave firing in CS-STA. Simulations demonstrate that CS-STA is capable of recovering the STA channel dataset with a smaller number of firings. The performance of CS-STA is evaluated in phantom experiments through comparisons with STA, multi-element STA, conventional focused mode and coherent plane wave imaging. The results demonstrate that, implemented with the same frame rate, CS-STA achieves higher or comparable resolution and contrast. Moreover, comparisons are conducted on the biceps brachii muscle and thyroid of a human subject, and the results demonstrate the feasibility and competitiveness of CS-STA in the in vivo conditions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据