4.4 Article

Research and Development of the High Stable Magnetic Field ReBCO Coil System Fundamental Technology for MRI

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TASC.2016.2645789

关键词

Superconducting magnet; magnetic resonance imaging (MRI); high-temperature superconducting (HTS); liquid helium free; ReBCO coil; field stability; field uniformity; imaging

资金

  1. Ministry of Economy, Trade and Industry (METI)
  2. Japan Agency for Medical Research and development (AMED)

向作者/读者索取更多资源

The superconducting magnet is effective to get a high stable and high magnetic field for magnetic resonance imaging (MRI). The current MRI superconducting magnet needed cooling in the liquid helium (4.2 K) to use NbTi superconducting wire. In the past few years, price increase and low availability of liquid helium has become a serious problem. Under such circumstances, the development of a high-temperature superconducting (HTS) coil dispensing with liquid helium cooling is greatly desired. The research and development project of the high stable magnetic field ReBCO coil system fundamental technology that started from the latter half of 2013 develops a ReBCO coil for 3 T MRI superconducting magnets. It gets a prospect of the practical use as the final aim. In this project, we will produce an HTS test coil of 300 mm bore experimentally and evaluate the magnetic field. This coil is cooled in less than 20 K by a GM refrigerator. We are going to make MRI used by the ReBCO coil field to evaluate the uniformity and stability of the magnetic field.