期刊
CHEMOSENSORS
卷 8, 期 4, 页码 -出版社
MDPI
DOI: 10.3390/chemosensors8040113
关键词
phenylalanine; sensor; Prussian Blue; cyclic voltammetry; electrochemistry
资金
- Dunarea de Jos University of Galati
This paper describes the sensitive properties of screen-printed carbon electrodes (SPCE) modified by using three different electroactive chemical compounds: Meldola's Blue, Cobalt Phthalocyanine and Prussian Blue, respectively. It was demonstrated that the Prussian Blue (PB) modified SPCE presented electrochemical signals with the highest performances in terms of electrochemical process kinetics and sensitivity in all the solutions analyzed. PB-SPCE was demonstrated to detect Phe through the influence it exerts on the redox processes of PB. The PB-SPCE calibration have shown a linearity range of 0.33-14.5 mu M, a detection limit (LOD) of 1.23 x 10(-8) M and the standard deviation relative to 3%. The PB-SPCE sensor was used to determine Phe by means of calibration and standard addition techniques on pure samples, on simple pharmaceutical samples or on multicomponent pharmaceutical samples. Direct determination of the concentration of 4 x 10(-6)-5 x 10(-5) M Phe in KCl solution showed that the analytical recovery falls in the range of 99.75-100.28%, and relative standard deviations in the range of 2.28-3.02%. The sensors were successfully applied to determine the Phe in pharmaceuticals. The validation of the method was performed by using the FTIR, and by comparing the results obtained by PB-SPCE in the analysis of three pharmaceutical products of different concentrations with those indicated by the producer.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐