3.8 Article

Fully-automated production of [68Ga]Ga-FAPI-46 for clinical application

期刊

出版社

SPRINGERNATURE
DOI: 10.1186/s41181-020-00112-x

关键词

Automated production; Imaging; FAP; FAPI-46; Fibroblast

资金

  1. Projekt DEAL

向作者/读者索取更多资源

Background[Ga-68]Ga-FAPI-46 is a promising radiopharmaceutical for in vivo detection of the fibroblast activation protein by positron emission tomography. Until now, the synthesis of [Ga-68]Ga-FAPI-46 has been only performed manually. Our aim was to evaluate the automated synthesis of this radiopharmaceutical on two different commercially available synthesis modules in order to make the tracer readily available for clinical application.ResultsThe synthesis of [Ga-68]Ga-FAPI-46 with different amounts of precursor (10-50 mu g) on the Modular Lab PharmTracer (MLPT) and Modular Lab eazy (ML eazy) from Eckert & Ziegler with a customized synthesis template and a customized single-use cassette yielded best results with 50 mu g FAPI-46 for clinical multi-dose application. All relevant quality control parameters tested (e.g. sterility, stability and radiochemical purity) were in accordance with the European Pharmacopoeia.Conclusions[Ga-68]Ga-FAPI-46 was successfully synthesized fully-automated on the synthesis modules Modular Lab PharmTracer and ML eazy and is, thus, available for multi-dose application in clinical settings.