3.8 Article

Multicycle Autoclave Decontamination of N95 Filtering Facepiece Respirators

期刊

APPLIED BIOSAFETY
卷 25, 期 3, 页码 150-156

出版社

MARY ANN LIEBERT, INC
DOI: 10.1177/1535676020924171

关键词

autoclave; decontamination; personal protective equipment; N95; surgical mask

资金

  1. Office of the Provost, University of Texas Medical Branch (UTMB)
  2. National Institutes of Health, National Institute of Allergy & Infectious Diseases National Biocontainment Laboratories (NBLs) Operations Support (UC7 Grant)

向作者/读者索取更多资源

Introduction: During pandemic situations like the one caused by the emergent coronavirus SARS-CoV-2, healthcare systems face the challenge of limited personal protective equipment and impaired supply chains. This problem poses a threat to healthcare workers, first responders, and the public, which demands solutions that can span the gap between institutional shortages and resupplies. Objectives: To examine the efficacy of autoclave-based decontamination for the reuse of single-use surgical masks and N95 filtering facepiece respirators (FFRs). This method is the most readily available form of decontamination in the hospital and laboratory settings. Methods: Three models of N95 FFRs and two procedural masks were evaluated in this study. A moist heat autoclave using four different autoclave cycles: 115 degrees C for one hour, 121.1 degrees C for 30 minutes, 130 degrees C for two minutes, and 130 degrees C for four minutes was used. After the autoclave process, the FFRs were NIOSH fit tested and particle counting was performed for both coarse particles of 5 micrometers (mu M) and fine particles from 0.1 mu M to 1.0 mu M. Results: We observed negligible alterations in the functionality and integrity of 3M 1805 and 3M 1870/1870+ N95 FFRs after three autoclave cycles. Surgical masks also showed minimal changes in functionality and integrity. The 3M 1860 FFR failed fit test after a single autoclave decontamination cycle. Discussion and Conclusion: The study finds that specific surgical masks and N95 FFR models can withstand autoclave decontamination for up to three cycles. Additionally, the autoclave cycles tested were those that could be readily achieved by both clinical and research institutions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据