4.6 Article

A microfluidic chip for on-line derivatization and application to in vivo neurochemical monitoring

期刊

ANALYST
卷 146, 期 3, 页码 825-834

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/d0an01729a

关键词

-

资金

  1. NIH [F32EB019800, R01EB003320]

向作者/读者索取更多资源

Microfluidic chips, combined with microdialysis, are used for benzoyl chloride derivatization to analyze small molecule neurochemicals in complex biological matrices. The chip reduces manual sample preparation steps and improves experimental throughput by conducting derivatization reactions on-line.
Microfluidic chips can perform a broad range of automated fluid manipulation operations for chemical analysis including on-line reactions. Derivatization reactions carried out on-chip reduce manual sample preparation and improve experimental throughput. In this work we develop a chip for on-line benzoyl chloride derivatization coupled to microdialysis, an in vivo sampling technique. Benzoyl chloride derivatization is useful for the analysis of small molecule neurochemicals in complex biological matrices using HPLC-MS/MS. The addition of one or more benzoyl groups to small, polar compounds containing amines, phenols, thiols, and certain alcohols improves reversed phase chromatographic retention, electrospray ionization efficiency, and analyte stability. The current derivatization protocol requires a three-step manual sample preparation, which ultimately limits the utility of this method for rapid sample collection and large sample sets. A glass microfluidic chip was developed for derivatizing microdialysis fractions on-line as they exit the probe for collection and off-line analysis with HPLC-MS/MS. Calibration curves for 21 neurochemicals prepared using the on-chip method showed linearity (R-2 > 0.99), limits of detection (0.1-500 nM), and peak area RSDs (4-14%) comparable to manual derivatization. Method temporal resolution was investigated both in vitro and in vivo showing rapid rise times for all analytes, which was limited by fraction length (3 min) rather than the device. The platform was applied to basal measurements in the striatum of awake rats where 19 of 21 neurochemicals were above the limit of detection. For a typical 2 h study, a minimum of 120 pipetting steps are eliminated per animal. Such a device provides a useful tool for the analysis of small molecules in biological matrices which may extend beyond microdialysis to other sampling techniques.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据