Current limits for flowmeter resistance in metabolic carts can negatively affect exercise performance

期刊

PHYSIOLOGICAL REPORTS
卷 9, 期 7, 页码 -

出版社

WILEY
DOI: 10.14814/phy2.14814

关键词

ergospirometry; exercise testing; flowmeter; resistance; respiratory muscle work; respiratory muscles

资金

  1. Swiss Federal Office of Sport [14-07]

向作者/读者索取更多资源

The study found that using a metabolic cart with airflow resistance in the upper range of guidelines significantly impacted exercise performance and respiratory pattern in young, healthy males.
Purpose: To investigate whether a metabolic cart using a flowmeter in the upper range of accepted resistance to airflow (<1.5 cmH(2)O.L-1.s(-1) for flows up to 14 L.s(-1), American Thoracic Society) negatively impacts exercise performance in healthy individuals. Methods: 16 recreationally active males (age 25 +/- 1 years, height 180 +/- 6 cm, weight 73.5 +/- 5.8 kg, all mean +/- SD) performed two incremental tests on a bicycle ergometer on each of two visits, using a metabolic cart with a flowmeter of either low (Oxycon Pro) or high (Innocor) airflow resistance. Mouth pressures, gas exchange, blood lactate concentration [La-], perception of breathlessness, respiratory, and leg exertion were assessed throughout the tests. Results: Tests performed with the Innocor were significantly shorter (15.3 +/- 3.2 vs. 15.8 +/- 3.3 min, p < 0.0001) and showed higher maximal flow resistance (1.3 +/- 0.2 vs. 0.3 +/- 0.0 cmH(2)O.L-1.s(-1), p < 0.0001). At end-exercise, peak oxygen consumption (-200 +/- 220 ml.min(-1), p < 0.0001), minute ventilation (-19.9 +/- 10.5 L.min(-1), p < 0.0001), breathing frequency (-5.4 +/- 5.2 breaths.min(-1), p < 0.0001), heart rate (-2.1 +/- 3.6 bpm, p = 0.002) and [La-] (-0.7 +/- 1.0 mmol.L-1, p < 0.0001), but not tidal volume (-0.1 +/- 0.2 L, p = 0.172) were lower with the Innocor, while the perception of breathlessness was higher (+3.8 +/- 5.1 points, p < 0.0001). Conclusions: Airflow resistance in the upper range of current guidelines can significantly affect exercise performance and respiratory pattern in young, healthy males during incremental exercise. The present results indicate the need to revisit guidelines for devices used in ergospirometry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据