Validation of plasma amino acid profile using UHPLC-mass spectrometer (QDa) as a screening method in a metabolic disorders reference centre: Performance and accreditation concerns

期刊

CLINICAL BIOCHEMISTRY
卷 92, 期 -, 页码 34-45

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2021.03.004

关键词

Liquid chromatography; Mass spectrometry; Derivatization; Aminoacids

向作者/读者索取更多资源

This study validated a rapid method for quantifying 27 amino acids in plasma, meeting French quality laboratory and SFEIM recommendations. The results were comparable to the standard method and can be used in routine practice.
Introduction: Amino acid (AA) analysis in plasma is essential for diagnosis and monitoring of inborn errors of metabolism (IEM). The efficacy of patient management is governed by the rapidity of AA profile availability, along with the robustness of the method. French quality guidelines and progress made in analytical techniques have led biologists to develop AA profile exploration via mass spectrometry (MS). Objectives: The aim of this study was to validate an analytical method with a single quadrupole mass spectrometer (MS) and to suggest reference values in regard to French quality and IEM society recommendations. Design and methods: Plasma samples from patients were deproteinised and derivatised with AccqTagTM reagent. Analysis was performed by reverse-phase chromatography coupled to QDA detector. We evaluated accuracy, intra-days and inter-days precision and limit of quantification by the beta-expectation tolerance interval method for 27 AA. Method comparison was performed with the standard method (ion exchange chromatography, IEC) on Jeol Aminotac (R) and to tandem MS. Reference values were established on AA concentrations of the cohort of patients who had no IEM. Results: Our method allowed the separations of almost all amino acids with a total run time of 12 min. Separation of isoleucine and alloisoleucine was incomplete (R = 0.55) but without impact on biological interpretation. Precision, accuracy and quantification were satisfactory (intra-days coefficient of variation (CV) was <5%, interdays precision CV <10% and accuracy <15%). The limits of quantification were validated between 1 and 900 & micro;mol/L. Results were comparable between the new method and IEC. Conclusion: Ultimately, we validated a rapid method on plasma for quantifying 27 amino acids that can be used in routine practice, according to French quality laboratories and SFEIM (French Society of Inborn Error of Metabolism) recommendations. Furthermore, estimated reference values were similar to those found in published studies focusing on other methods. Despite a lower specificity compared to tandem MS, the simplicity and rapidity of our method are the main advantage of this technique and place it as a major tool in IEM diagnosis and management.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据