3.9 Article

METHOD VALIDATION FOR THE DETERMINATION OF EXCHANGEABLE CATIONS IN NATURAL ZEOLITES USING INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETRY

期刊

STUDIA UNIVERSITATIS BABES-BOLYAI CHEMIA
卷 66, 期 2, 页码 81-94

出版社

UNIV BABES-BOLYAI
DOI: 10.24193/subbchem.2021.2.07

关键词

zeolites; exchangeable cations; ICP-OES; method validation; cation exchange capacity

资金

  1. Competitiveness Operational Programme of the Ministry of European Funds [7/01.09.2016, MySMIS 105654]

向作者/读者索取更多资源

This paper presents the validation of a method for determining exchangeable cations in zeolites using ICP-OES, showing rapid, accurate analysis. The method validation demonstrates that the method has suitable precision and accuracy, meeting the requirements of the ISO/IEC 17025:2017 standard.
The paper presents the validation of a method for the determination of exchangeable cations (K+, Na+, Ca2+ and Mg2+) in zeolites by inductively coupled plasma optical emission spectrometry (ICP-OES), after extraction in ammonium chloride solution. Amongst the analytical techniques that can be used for exchangeable cations measurement, ICP-OES provides rapid, robust, multi-element analysis on a wide range of concentrations. The exchange capacity is one of the most important parameters of zeolites since it indicates their adsorption capacity in different processes. Method validation is an essential requirement for testing laboratories in order to provide trustworthy results in accordance with ISO/IEC 17025:2017 standard. The main figures of merit were studied, and the measurement uncertainty was assessed. The selectivity assay showed no significant non-spectral matrix effect. The linearity study was conducted for the calibration curves in the range of 0.02-20 mg L-1 for each exchangeable cation. Limits of quantification were 0.005 mEq 100 g(-1) for Na+, 0.011 mEq 100 g(-1) for Mg2+, 0.002 mEq 100 g(-1) for K+, 0.003 mEq 100 g(-1) for Ca2+. Relative standard deviations of repeatability (RSDr) (n=6 parallel samples) were 4.50 % for Na+, 4.43 % for Mg2+, 6.55 % for K+ and 5.53 % for Ca2+. Recoveries (%) estimated using a certified reference material (CRM BCS-CRM 375/1), for total content of cation oxides were in the range 92-103 %. Fourteen zeolite samples from Racos deposit, Romania were analysed, and according to the chemical composition and Xray diffraction the main mineral in zeolitic tuffs is clinoptilolite-type. The obtained figures of merit demonstrate that the method has a suitable level of precision and accuracy for the intended purpose.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据