资金

  1. German Federal Ministry of Education and Research (BMBF, project Nav EVAR) [13GW0228]
  2. Ministry of Economic Affairs, Employment, Transport and Technology of Schleswig-Holstein

向作者/读者索取更多资源

A catheter prototype using a multicore fiber and three EM sensors for shape sensing and localization was developed for EVAR procedures. Evaluation showed average errors ranging from 0.99 to 2.29 mm and maximum errors from 1.73 to 2.99 mm in different shapes, indicating the feasibility and promise of accurate shape localization with this method. Future work will focus on further development for catheter guidance in EVAR procedures.
Currently, fluoroscopy and conventional digital subtraction angiography are used for imaging guidance in endovascular aortic repair (EVAR) procedures. Drawbacks of these image modalities are X-ray exposure, the usage of contrast agents and the lack of depth information. To overcome these disadvantages, a catheter prototype containing a multicore fiber with fiber Bragg gratings for shape sensing and three electromagnetic (EM) sensors for locating the shape was built in this study. Furthermore, a model for processing the input data from the tracking systems to obtain the located 3D shape of the first 38 cm of the catheter was introduced: A spatial calibration between the optical fiber and each EM sensor was made in a calibration step and used to obtain the located shape of the catheter in subsequent experiments. The evaluation of our shape localization method with the catheter prototype in different shapes resulted in average errors from 0.99 to 2.29 mm and maximum errors from 1.73 to 2.99 mm. The experiments showed that an accurate shape localization with a multicore fiber and three EM sensors is possible, and that this catheter guidance is promising for EVAR procedures. Future work will be focused on the development of catheter guidance based on shape sensing with a multicore fiber, and the orientation and position of less than three EM sensors.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据