3.8 Proceedings Paper

Non-invasive Thermometry During Hyperthermia Using Differential Microwave Imaging Approach

出版社

IEEE

关键词

Microwave Imaging; Thermometry; Hyperthermia; Born Approximation (BA); TSVD; Dielectric Properties

资金

  1. COST (European Cooperation in Science and Technology) [CA17115]
  2. Ministry of Education, Youth and Sports of the Czech Republic [LTC19031]

向作者/读者索取更多资源

The evaluation of non-invasive microwave thermometry using a differential microwave imaging approach was conducted in an experimental setup, achieving a promising agreement between measured and reconstructed electrical conductivity. Specific areas for future improvement of microwave thermometry were identified through this study.
The evaluation of non-invasive microwave thermometry using a differential microwave imaging approach is investigated in an experimental setup. The setup consists of a monopole microwave tomographic system, a muscle phantom, and a circulation/heating system. The saline solution is used as the heated medium inside the muscle phantom. Based on Born Approximation and regularization by TSVD we reconstructed the distribution of both dielectric parameters, relative permittivity and electrical conductivity inside the investigation area. A promising agreement between measured and reconstructed electrical conductivity was achieved and some specifics for future improvement of microwave thermometry were identified.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据