4.5 Article

Radiation dose reduction during transjugular intrahepatic portosystemic shunt implantation using a new imaging technology

期刊

EUROPEAN JOURNAL OF RADIOLOGY
卷 86, 期 -, 页码 284-288

出版社

ELSEVIER IRELAND LTD
DOI: 10.1016/j.ejrad.2016.11.028

关键词

Transjugular intrahepatic portosystemic shunt (TIPS); Radiation dosage; Image-processing technology; Interventional radiology; Liver

向作者/读者索取更多资源

Objective: To compare patient radiation dose in patients undergoing transjugular intrahepatic portosystemic shunt (TIPS) implantation before and after an imaging-processing technology upgrade. Methods: In our retrospective single-center-study, cumulative air kerma (AK), cumulative dose area product (DAP), total fluoroscopy time and contrast agent were collected from an age- and BMI-matched collective of 108 patients undergoing TIPS implantation. 54 procedures were performed before and 54 after the technology upgrade. Mean values were calculated and compared using two-tailed t-tests. Two blinded, independent readers assessed DSA image quality using a four-rank likert scale and the Wilcoxcon test. Results: The new technology demonstrated a significant reduction of 57% of mean DAP (402.8 vs. 173.3 Gycm(2), p < 0.001) and a significant reduction of 58% of mean AK (1.7 vs. 0.7 Gy, p < 0.001) compared to the precursor technology. Time of fluoroscopy (26.4 vs. 27.8 min, p = 0.45) and amount of contrast agent (109.4 vs. 114.9m1, p = 0.62) did not differ significantly between the two groups. The DSA image quality of the new technology was not inferior (2.66 vs. 2.77, p = 0.56). Conclusions: In our study the new imaging technology halved radiation dose in patients undergoing TIPS maintaining sufficient image quality without a significant increase in radiation time or contrast consumption. (C) 2016 Elsevier Ireland Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据