期刊
EUROPEAN JOURNAL OF PEDIATRICS
卷 176, 期 5, 页码 573-580出版社
SPRINGER
DOI: 10.1007/s00431-017-2874-3
关键词
Validation; Blood pressure measuring device; Children; ESH; Custo screen pediatric
类别
资金
- Custo Med GmbH, Ottobrunn, Germany
The purpose of the study was to validate the ambulatory blood pressure monitoring (ABPM) device custo screen pediatric in children aged 3 to 12 years according to the International Protocol of the European Society of Hypertension (ESH-IP revision 2010). Thirty-three children were included and systolic and diastolic blood pressure measurements were performed according to the ESH-IP. The protocol was modified for children considering data from the German Health Interview and Examination Survey for Children and Adolescents (KIGGS). The custo screen pediatric met all the requirements of the ESH-IP. The mean difference between the test device and the reference was -1.4 +/- 3.0 mmHg for systolic blood pressure (SBP) and -0.7 +/- 3.2 mmHg for diastolic blood pressure (DBP). For SBP and DBP, all 99 measurements were within the absolute difference of 10 mmHg between the test device and the reference. As to part 2 of the protocol, for DBP in all subjects, two out of three measurements were within 5 mmHg between the device and the standard, whereas for SBP in 32 of 33 subjects, two out of three measurements were within this range. Conclusion: The custo screen pediatric met all criteria of the ESH-IP review 2010, modified for children from 3 to about 12 years, and can be recommended for ABPM in children.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。