Accurate templating for arthroplasty procedures relies on reliably estimating image magnification. The authors' simple device using staggered marker balls in orthogonal imaging proved to be more precise compared to assuming a constant 20% magnification for all x-rays, with the novel technique having no outliers exceeding 4% error.
Accurate templating for arthroplasty procedures depends on reliably estimating image magnification. The authors designed a simple device and technique to accomplish this. It is a wedge with four 25-mm balls staggered at different heights, each representing a potential marker. The proper ball to use as the scaling marker is identified on an orthogonal view on the basis of its relationship with the anatomic feature of interest. Reviewing x-rays with known acetabular component diameters, the authors found that the staggered balls technique had a scaling error of 1.48 +/- 0.89%. Assuming a constant 20% magnification for all x-rays resulted in an error of 1.65 +/- 1.59%. Although similarly accurate, the difference in precision was found to be significant (P=0.012). The constant magnification method resulted in 2 outliers with >4% error, whereas there were no outliers with the novel technique. Staggered marker balls combined with orthogonal imaging is accurate and improves the precision of estimating radiograph magnification.