4.5 Article

Transcatheter Heart Valve Implantation in Bicuspid Patients with Self-Expanding Device

期刊

BIOENGINEERING-BASEL
卷 8, 期 7, 页码 -

出版社

MDPI
DOI: 10.3390/bioengineering8070091

关键词

transcatheter aortic valve implantation; bicuspid aortic valve; finite-element analysis; fluid-solid interaction

向作者/读者索取更多资源

This study conducted numerical simulations of TAVI in patients with bicuspid aortic valve (BAV) to evaluate the feasibility and risks of transcatheter heart valve (THV) implantation. Through structural mechanics and fluid-solid interaction analysis, it was found that THV exhibited slight asymmetric expansion and potential paravalvular leakage risk in BAV anatomy.
Bicuspid aortic valve (BAV) patients are conventionally not treated by transcathether aortic valve implantation (TAVI) because of anatomic constraint with unfavorable outcome. Patient-specific numerical simulation of TAVI in BAV may predict important clinical insights to assess the conformability of the transcathether heart valves (THV) implanted on the aortic root of members of this challenging patient population. We aimed to develop a computational approach and virtually simulate TAVI in a group of n.6 stenotic BAV patients using the self-expanding Evolut Pro THV. Specifically, the structural mechanics were evaluated by a finite-element model to estimate the deformed THV configuration in the oval bicuspid anatomy. Then, a fluid-solid interaction analysis based on the smoothed-particle hydrodynamics (SPH) technique was adopted to quantify the blood-flow patterns as well as the regions at high risk of paravalvular leakage (PVL). Simulations demonstrated a slight asymmetric and elliptical expansion of the THV stent frame in the BAV anatomy. The contact pressure between the luminal aortic root surface and the THV stent frame was determined to quantify the device anchoring force at the level of the aortic annulus and mid-ascending aorta. At late diastole, PVL was found in the gap between the aortic wall and THV stent frame. Though the modeling framework was not validated by clinical data, this study could be considered a further step towards the use of numerical simulations for the assessment of TAVI in BAV, aiming at understanding patients not suitable for device implantation on an anatomic basis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据