4.7 Article

Real-Time Optimization of Retinal Ganglion Cell Spatial Activity in Response to Epiretinal Stimulation

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNSRE.2021.3138297

关键词

Retina; Shape; Electrodes; Optimization; Calcium; Implants; Ganglia; Calcium imaging; closed-loop optimization; electrical stimulation; retinal ganglion cell; retinal prostheses

资金

  1. Research toPrevent Blindness
  2. National Institutes of Health (NIH) [EY013934]

向作者/读者索取更多资源

Retinal prostheses aim to improve visual perception in patients with photoreceptor degeneration-induced blindness, but the current limitations in shape and letter perception are due to low spatial resolution. Research has shown that better patient outcomes are correlated with spatially separate phosphenes. By using an optimization algorithm to find optimal stimulation parameters for focal RGC activation, the time required for optimizing implant operation can be reduced.
Retinal prostheses aim to improve visual perception in patients blinded by photoreceptor degeneration. However, shape and letter perception with these devices is currently limited due to low spatial resolution. Previous research has shown the retinal ganglion cell (RGC) spatial activity and phosphene shapes can vary due to the complexity of retina structure and electrode-retina interactions. Visual percepts elicited by single electrodes differ in size and shapes for different electrodes within the same subject, resulting in interference between phosphenes and an unclear image. Prior work has shown that better patient outcomes correlate with spatially separate phosphenes. In this study we use calcium imaging, in vitro retina, neural networks (NN), and an optimization algorithm to demonstrate a method to iteratively search for optimal stimulation parameters that create focal RGC activation. Our findings indicate that we can converge to stimulation parameters that result in focal RGC activation by sampling less than 1/3 of the parameter space. A similar process implemented clinically can reduce time required for optimizing implant operation and enable personalized fitting of retinal prostheses.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据