ANALYTICAL & BIOANALYTICAL ELECTROCHEMISTRY
卷 13, 期 3, 页码 383-392出版社
UNIV TEHRAN, FAC CHEMISTRY, CENTER EXCELLENCE ELECTROCHEMISTRY
关键词
Ion selective electrode; PVC Membrane; Dithizone; Pb2+; Nernstian slope
The successful preparation and characterization of Pb2+ Ion Selective Electrode using dithizone as ion carrier was achieved, showing good performance within a certain concentration range and under optimized working conditions.
The research on the preparation and characterization of Pb2+ Ion Selective Electrode (ISE) using dithizone (DTZ) as ion carrier has been done. Membrane of ISE was made with the optimum composition with ratio of DTZ : PVC : DOP was 6 : 3 : 1. The membrane was doped using 1M Pb2+ solution within 7 days, and characterized using FT-IR to determine the shift of wave number from the spectrum of the group from membrane before and after doped. The results of characterization of FT-IR spectra of membrane after doped, shows that there is a shift of wave number from absorption spectra of secondary amine (N-H) group and absorption spectra of thiol (S-H) group didn't appear. The result of characterization of Pb2+ ISE shows that ISE worked well with range concentration of 1x10(-4)-1x10(-1) M, with Nernst slope was 29.5 mV/decade and the limit of detection was 1.5x10(-4) M. The electrode worked optimally at pH 5, the response time was 9-48 seconds and can be used within 6 weeks. The presence of interfere ions, such as Ag+, Zn2+, Cu2+, Cd2+, Cl- and SO42- didn't affect to the determination of Pb2+ ions in the samples.