Prospective validation of smartphone-based heart rate and respiratory rate measurement algorithms

期刊

COMMUNICATIONS MEDICINE
卷 2, 期 1, 页码 -

出版社

SPRINGERNATURE
DOI: 10.1038/s43856-022-00102-x

关键词

-

资金

  1. Google Health

向作者/读者索取更多资源

This study validates the accuracy of smartphone camera-based techniques in measuring heart rate and respiratory rate across various pre-defined subgroups.
Background Measuring vital signs plays a key role in both patient care and wellness, but can be challenging outside of medical settings due to the lack of specialized equipment. Methods In this study, we prospectively evaluated smartphone camera-based techniques for measuring heart rate (HR) and respiratory rate (RR) for consumer wellness use. HR was measured by placing the finger over the rear-facing camera, while RR was measured via a video of the participants sitting still in front of the front-facing camera. Results In the HR study of 95 participants (with a protocol that included both measurements at rest and post exercise), the mean absolute percent error (MAPE) +/- standard deviation of the measurement was 1.6% +/- 4.3%, which was significantly lower than the pre-specified goal of 5%. No significant differences in the MAPE were present across colorimeter-measured skin-tone subgroups: 1.8% +/- 4.5% for very light to intermediate, 1.3% +/- 3.3% for tan and brown, and 1.8% +/- 4.9% for dark. In the RR study of 50 participants, the mean absolute error (MAE) was 0.78 +/- 0.61 breaths/min, which was significantly lower than the pre-specified goal of 3 breaths/min. The MAE was low in both healthy participants (0.70 +/- 0.67 breaths/min), and participants with chronic respiratory conditions (0.80 +/- 0.60 breaths/min). Conclusions These results validate the accuracy of our smartphone camera-based techniques to measure HR and RR across a range of pre-defined subgroups.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据