3.8 Article

Thalamic local field potentials recorded using the deep brain stimulation pulse generator

期刊

CLINICAL NEUROPHYSIOLOGY PRACTICE
卷 7, 期 -, 页码 103-106

出版社

ELSEVIER
DOI: 10.1016/j.cnp.2022.03.002

关键词

Tremor; Adaptive; Closed-loop; Deep brain stimulation; Essential tremor

向作者/读者索取更多资源

This article reports a case of recording thalamic LFP activity using a commercially available sensing enabled DBS pulse generator, revealing peak tremor frequency during tremor evoking movements and spectrum characteristics of LFP signal. Coherence analysis demonstrated consistency between peripherally recorded tremor and LFP signal, providing important information for further evaluation of the clinical potential of fully implantable systems.
Background: Essential tremor (ET) is one of the most common movement disorders, and continuous deep brain stimulation (DBS) is an established treatment for medication-refractory cases. However, the need for increasing stimulation intensities, with unpleasant side effects, and DBS tolerance over time can be problematic. The advent of novel DBS devices now provides the opportunity to longitudinally record LFPs using the implanted pulse generator, which opens up possibilities to implement adaptive DBS algorithms in a real-life setting.Methods: Here we report a case of thalamic LFP activity recorded using a commercially available sensing enabled DBS pulse generator (Medtronic Percept PC).Results: In the OFF-stimulation condition, a peak tremor frequency of 3.8 Hz was identified during tremor evoking movements as assessed by video and accelerometers. Activity at the same and supraharmonic frequency was seen in the frequency spectrum of the LFP data from the left vim nucleus during motor tasks. Coherence analysis showed that peripherally recorded tremor was coherent with the LFP signal at the tremor frequency and supraharmonic frequency.Conclusion: This is the first report of recorded tremor-related thalamic activity using the electrodes and pulse generator of an implanted DBS system. Larger studies are needed to evaluate the clinical potential of these fully implantable systems, and ultimately pulse generators with sensing-coupled algorithms driving stimulation, to really close the loop.(c) 2022 International Federation of Clinical Neurophysiology. Published by Elsevier B.V. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据