4.4 Article

Importance of oxide film in endovascular biodegradable zinc stents

期刊

SURFACE INNOVATIONS
卷 4, 期 3, 页码 133-140

出版社

ICE PUBLISHING
DOI: 10.1680/jsuin.16.00014

关键词

biocompatibility; biodegradable; biointerface

资金

  1. US National Institutes of Health-National Institute of Biomedical Imaging and Bioengineering [5R21 EB 019118-02]
  2. US National Institutes of Health-National Heart, Lung, and Blood Institute [1R15 HL129199-01]
  3. Michigan Initiative for Innovation and Entrepreneurship's Technology Commercialization program

向作者/读者索取更多资源

A better understanding of the surface oxide film on biodegradable metals could lead to the design of smarter, surface-responsive biodegradable stents with good biocompatibility and controlled degradation rates at different stages of implantation. In this study, the surface finish of zinc was manipulated through different material processing conditions, including oxidation, electropolishing and anodization. Implant wires with oxide films of varying surface characteristics were degraded in a living organism by using a rodent model to understand their response to a biological endovascular environment. It was found that the degradation rate was mainly dependent on the quality and stability of oxide film. Defects/cracks in the oxide film structure appear to serve as local corrosion sites, and their increased density accelerates the biocorrosion rate.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据