Computational Contact Pressure Prediction of CoCrMo, SS 316L and Ti6Al4V Femoral Head against UHMWPE Acetabular Cup under Gait Cycle

期刊

JOURNAL OF FUNCTIONAL BIOMATERIALS
卷 13, 期 2, 页码 -

出版社

MDPI
DOI: 10.3390/jfb13020064

关键词

CoCrMo; contact pressure; SS 316L; Ti6Al4V; total hip arthroplasty; UHMWPE

资金

  1. World Class Research UNDIP [118-23/UN7.6.1/PP/2021]
  2. DIPA of Public Service Agency of Sriwijaya University 2021 [0014/UN9/SK.LP2M.PT/2021]

向作者/读者索取更多资源

This study computationally evaluated the contact pressure between UHMWPE acetabular cup and metal femoral head, and found that Ti6Al4V-on-UHMWPE can reduce cumulative contact pressure.
Due to various concerns about the use of metal-on-metal that is detrimental to users, the use of metal as acetabular cup material was later changed to ultra high molecular weight polyethylene (UHMWPE). However, the wear on UHMWPE releases polyethylene wear particles, which can trigger a negative body response and contribute to osteolysis. For reducing the wear of polyethylene, one of the efforts is to investigate the selection of metal materials. Cobalt chromium molybdenum (CoCrMo), stainless steel 316L (SS 316L), and titanium alloy (Ti6Al4V) are the frequently employed materials. The computational evaluation of contact pressure was carried out using a two-dimensional axisymmetric model for UHMWPE acetabular cup paired with metal femoral head under gait cycle in this study. The results show Ti6Al4V-on-UHMWPE is able to reduce cumulative contact pressure compared to SS 316L-on-UHMWPE and CoCrMo-on-UHMWPE. Compared to Ti6Al4V-on-UHMWPE at peak loading, the difference in cumulative contact pressure to respective maximum contact pressure is 9.740% for SS 316L-on-UHMWPE and 11.038% for CoCrMo-on-UHMWPE.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据