4.6 Article

3D printed microfluidic Coulter counter for blood cell analysis

期刊

ANALYST
卷 147, 期 14, 页码 3225-3233

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/d2an00633b

关键词

-

向作者/读者索取更多资源

This article introduces a disposable Coulter counting analysis kit based on 3D microfluidic printing technology, which improves the accuracy of Coulter cell count analysis and solves the problem of clogging. The kit is convenient, hygienic, and can be widely used in POCT detection scenarios.
Coulter counters are ubiquitous in everyday life; however, with reduced orifice size shrinkage, there is an increased risk of clogging. Herein, a 3D microfluidic printing-based single-use kit is presented for analyzing biological samples and performing accurate Coulter cell count analysis (e.g., white blood cells in the blood). The gem hole eliminates the traditional design concept of integration inside the detection instrument, innovatively causing it to be independent from the analysis instrument. Further, integrating the hole in the analysis box enables the design of a separate detection module. The analysis box is disposable, convenient, and hygienic; avoids cross-infection; solves the problem of clogging of tiny holes from a new perspective; and no longer requires uneconomical and inconvenient methods, such as flushing, cauterization, and fluid focusing, through solid water flow. Further development of the newly designed 3D printing analysis box can enable its extensive use in POCT (point-of-care) detection scenarios. Moreover, through mass production, the issue of cost will be eliminated.