Low Voltage and Low Power Front Panel Design for 12 Lead ECG

期刊

IEEE ACCESS
卷 10, 期 -, 页码 69455-69461

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/ACCESS.2022.3184782

关键词

Electrocardiography; Multiplexing; Threshold voltage; Substrates; MOSFET; Switches; Logic gates; Analog multiplexer; DTMOS; ECG; data acquisition system; noise analysis; signal transmission

向作者/读者索取更多资源

With the increasing use of portable devices, Electro Cardiogram (ECG) plays a vital role in the medical field as an important emergency medical portable system. However, the excessive contact impedance of electrodes leads to an increase in power dissipation in the data acquisition system. This paper proposes a CMOS 12 lead ECG data acquisition circuit based on analog multiplexers to address this issue.
The use of portable devices is increasing day by day. Mainly these devices play a vital role in medical field when the doctors are unavailable. In that one of the important emergency medical portable systems is Electro Cardiogram (ECG). ECGs are used to measure the abnormalities of the heart by performing electrical activities. It can be done by attaching electrodes on the body for few minutes. To collect the information from electrodes data acquisition system is used in ECGs or in any biomedical systems. The major challenge in implementing data acquisition system is the increase in power dissipation due to the excessive contact impedance of electrodes. As a solution to this, A CMOS 12 lead ECG data acquisition circuit is designed by using analog multiplexers is proposed in this paper. The proposed system is designed by using improved DTMOS based analog multiplexers and these works with a power supply of +/- 0.2 V with a dynamic voltage range from 1uv to 200mv with the Power dissipation of 140 nW.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据