4.7 Article

Graphene-Enabled Electrodes for Electrocardiogram Monitoring

期刊

NANOMATERIALS
卷 6, 期 9, 页码 -

出版社

MDPI
DOI: 10.3390/nano6090156

关键词

graphene; electrocardiogram (ECG); long-term monitoring; dry electrodes; contact impedance; Raman spectroscopy

资金

  1. Ministry of National Education of Turkey
  2. research group of Doclab in Brunel University London
  3. 2-DTech company in Manchester

向作者/读者索取更多资源

The unique parameters of graphene (GN)-notably its considerable electron mobility, high surface area, and electrical conductivity-are bringing extensive attention into the wearable technologies. This work presents a novel graphene-based electrode for acquisition of electrocardiogram (ECG). The proposed electrode was fabricated by coating GN on top of a metallic layer of a Ag/AgCl electrode using a chemical vapour deposition (CVD) technique. To investigate the performance of the fabricated GN-based electrode, two types of electrodes were fabricated with different sizes to conduct the signal qualities and the skin-electrode contact impedance measurements. Performances of the GN-enabled electrodes were compared to the conventional Ag/AgCl electrodes in terms of ECG signal quality, skin-electrode contact impedance, signal-to-noise ratio (SNR), and response time. Experimental results showed the proposed GN-based electrodes produced better ECG signals, higher SNR (improved by 8%), and lower contact impedance (improved by 78%) values than conventional ECG electrodes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据