4.2 Article

Comparing the Performance of Popular MEG/EEG Artifact Correction Methods in an Evoked-Response Study

期刊

出版社

HINDAWI LTD
DOI: 10.1155/2016/7489108

关键词

-

向作者/读者索取更多资源

We here compared results achieved by applying popular methods for reducing artifacts in magnetoencephalography (MEG) and electroencephalography (EEG) recordings of the auditory evoked Mismatch Negativity (MMN) responses in healthy adult subjects. We compared the Signal Space Separation (SSS) and temporal SSS (tSSS) methods for reducing noise from external and nearby sources. Our results showed that tSSS reduces the interference level more reliably than plain SSS, particularly for MEG gradiometers, also for healthy subjects not wearing strongly interfering magnetic material. Therefore, tSSS is recommended over SSS. Furthermore, we found that better artifact correction is achieved by applying Independent Component Analysis (ICA) in comparison to Signal Space Projection (SSP). Although SSP reduces the baseline noise level more than ICA, SSP also significantly reduces the signal-slightly more than it reduces the artifacts interfering with the signal. However, ICA also adds noise, or correction errors, to the wave form when the signal-to-noise ratio (SNR) in the original data is relatively low-in particular to EEG and to MEG magnetometer data. In conclusion, ICA is recommended over SSP, but one should be careful when applying ICA to reduce artifacts on neurophysiological data with relatively low SNR.