4.4 Article

Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices during Thyroid and Parathyroid Surgery in a Porcine Model

期刊

出版社

JOURNAL OF VISUALIZED EXPERIMENTS
DOI: 10.3791/63732

关键词

-

资金

  1. Kaohsiung Medical University Hospital, Kaohsiung Medical University [KMUH109-9M44]
  2. Kaohsiung Municipal Siaogang Hospital/Kaohsiung Medical University Research Center [KMHK-DK(C)110009, I-109-04, H-109-05, I-108-02]
  3. Ministry of Science and Technology, Taiwan [MOST 109-2628-B-037-014, MOST 110-2314-B-037-104-MY2, MOST 110-2314-B-037-120]

向作者/读者索取更多资源

This article introduces experimental studies using a porcine model to investigate the safe activation and cooling parameters of surgical energy devices (SEDs) in thyroid and parathyroid surgery. The studies use electrophysiological and thermographic procedures to evaluate the safe activation distance and cooling time, ensuring the avoidance of recurrent laryngeal nerve injury.
In thyroid and parathyroid surgery, surgical energy devices (SEDs) provide more efficient hemostasis than conventional clamp-and-tie hemostasis in areas with rich blood supply. However, when a SED is activated near the recurrent laryngeal nerve (RLN), the heat generated by the SED may injure the nerve irreversibly. To safely apply SEDs in thyroid/parathyroid surgery, this article introduces experimental porcine model studies to investigate the activation and cooling safety parameters of SEDs in standardized electrophysiological (EP) and thermographic (TG) procedures, respectively. In the EP safety parameter experiments, continuous intraoperative neuromonitoring (C-IONM) is applied to demonstrate the RLN function in real-time. The EP activation study evaluates the safe activation distance of SEDs; the EP cooling study evaluates the safe cooling time of SEDs. In the TG safety parameter experiment, a thermal imaging camera is used to record the temperature change after activating the SED. The TG activation study evaluates the lateral thermal spread distance after SED activation in a dry or humid environment and whether smoke and splashing are generated; the TG cooling study evaluates the cooling time. This will help establish the safety parameters of newly developed SEDs used in thyroid/parathyroid surgery and provide safety guidelines to avoid RLN injury and related complications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据