期刊
TALANTA
卷 266, 期 -, 页码 -出版社
ELSEVIER
DOI: 10.1016/j.talanta.2023.125100
关键词
Electrochemical sensor; Molecularly imprinted polymer; MXene fibers; Hydrocortisone
In this study, a molecularly imprinted electrochemical sensor based on MXene fibers was proposed. MXene fibers were prepared using the wet spinning technique, and molecularly imprinted polymers were synthesized on the surface of MXene fibers using electro-polymerization. The electrochemical sensor exhibited superior selectivity and sensitivity for hydrocortisone, and showed excellent application performance in serum.
A molecularly imprinted electrochemical sensor based on MXene fibers was proposed in this work. Firstly, the wet spinning technique prepared MXene fibers with a large aspect ratio, which can make the sheet-like MXene uniformly arranged, avoiding the agglomeration of MXene and improving the electrical conductivity. Afterwards, molecularly imprinted polymers (MIPs) with specific recognition sites were synthesized on the surface of MXene fibers using the electro-polymerization method. The electrochemical sensor utilized the advantages of MXene fibers and molecular imprinting techniques to gain superior selectivity and sensitivity of hydrocortisone (HC). Electrochemical tests with different concentrations of HC (0.5 nM-10.0 mu M) under optimal measurement conditions exhibited excellent linearity and a limit of detection (LOD) of 0.17 nM. Furthermore, the electrochemical sensor displayed excellent selectivity, interference resistance, reproducibility, stability and outstanding application performance in serum. This work has promising applications in trace analysis in real sample.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据