4.7 Article

Effects on Retinal Stimulation of the Geometry and the Insertion Location of Penetrating Electrodes

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNSRE.2023.3317496

关键词

Retina; Electrodes; Conductivity; Electric potential; Shape; Photoreceptors; Microelectrodes; Artificial vision; retinal implant; retinal prosthesis; computational simulation; penetrating electrodes

向作者/读者索取更多资源

This study used the finite element method to simulate the electric fields generated by three different designs of 3D microelectrodes in a retina model with a stratified conductivity profile. The results showed that a pillar electrode with an insulating wall generated a stronger and more confined electric field, which is preferred for high-resolution retinal prostheses. Additionally, the insertion depth of the electrode was found to significantly affect the magnitude and shape of the potential distribution.
Retinal implants have been developed and implanted to restore vision from outer retinal degeneration, but their performance is still limited due to the poor spatial resolution. To improve the localization of stimulation, microelectrodes in various three-dimensional (3D) shapes have been investigated. In particular, computational simulation is crucial for optimizing the performance of a novel microelectrode design before actual fabrication. However, most previous studies have assumed a uniform conductivity for the entire retina without testing the effect of electrodes placement in different layers. In this study, we used the finite element method to simulate electric fields created by 3D microelectrodes of three different designs in a retina model with a stratified conductivity profile. The three electrode designs included two conventional shapes - a conical electrode (CE) and a pillar electrode (PE); we also proposed a novel structure of pillar electrode with an insulating wall (PEIW). A quantitative comparison of these designs shows the PEIW generates a stronger and more confined electric field with the same current injection, which is preferred for high-resolution retinal prostheses. Moreover, our results demonstrate both the magnitude and the shape of potential distribution generated by a penetrating electrode depend not only on the geometry, but also substantially on the insertion depth of the electrode. Although epiretinal insertions are mainly discussed, we also compared results for subretinal insertions. The results provide valuable insights for improving the spatial resolution of retinal implants using 3D penetrating microelectrodes and highlight the importance of considering the heterogeneity of conductivities in the retina.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据