期刊
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
卷 72, 期 -, 页码 -出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TIM.2023.3311056
关键词
Cramer-Rao lower bound (CRLB); displacement measurement; incoherent optical method; sensitivity limit
This study investigates the fundamental sensitivity limit of optical displacement measurement methods by deriving the Cramer-Rao lower bound, which represents the minimum standard deviation of measurable displacement.
The sensitivity (i.e., the minimum measurable displacement) of incoherent optical methods for full-field displacement measurements [e.g., optical flow (OF) and digital image correlation (DIC)] is essentially limited by the bit depth of the digital camera due to the quantization with round-off errors. The sensitivity limit has been recently exceeded by a multipixel averaging method, achieving super-sensitivity. However, a critical question has not been answered: Is there a limit to the achieved super-sensitivity? In this work, we investigate the fundamental sensitivity limit regardless of the involved signal-processing algorithms. Specifically, we derive the Cramer-Rao lower bound (CRLB) on the variance of displacement estimators, that is, the minimum value of the variance that may be attained by any unbiased estimator. Since the minimum standard deviation (SD) of the displacement estimation corresponds to the minimum measurable displacement, the CRLB is suitable to represent the fundamental sensitivity limit. Furthermore, the derived fundamental sensitivity limit is illustrated and validated by theoretical simulations and laboratory experiments, respectively.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐