Enhancing Therapeutic Drug Monitoring in Inflammatory Bowel Disease: A Comparative Analysis of Rapid Point-of-Care Infliximab, Adalimumab and Anti-Drug Antibodies' Determination against ELISA

期刊

PHARMACEUTICS
卷 15, 期 11, 页码 -

出版社

MDPI
DOI: 10.3390/pharmaceutics15112615

关键词

point-of-care; infliximab; adalimumab; anti-drug antibodies; ELISA

向作者/读者索取更多资源

This study compared the performance of the point-of-care (POC) test Quantum blue with the traditional ELISA assay in patients with inflammatory disease. The results showed higher values for infliximab (IFX) using the POC test, but no differences in adalimumab (ADL) levels were observed. There was a good correlation between the two assays, but significant differences were found in classifying IFX samples between therapeutic and supra-therapeutic ranges.
The introduction of point-of-care (POC) assays into clinical practice in patients with inflammatory disease enables on-demand therapeutic decision making. The aim of this study was to compare the POC test Quantum blue (B & uuml;hlmann Laboratories) for infliximab (IFX), adalimumab (ADL), and its anti-drug antibodies with the traditional ELISA assay (Promonitor). A total of 200 serum samples were analyzed. Samples were classified into the following three different groups; sub-therapeutic range (IFX < 3 mu g/mL and ADL < 5 mu g/mL); therapeutic range (IFX: 3-7 mu g/mL and ADL: 5-12 mu g/mL) and supra-therapeutic range (IFX levels > 7 mu g/mL and ADL levels > 12 mu g/mL). Significant higher values were measured using the POC test (p < 0.001) for IFX results but no differences in ADL trough levels were observed (p = 0.3101). Spearman's correlation indicated a good correlation between the two assays (rs = 0.88 for ADL and rs = 0.93 for IFX), and McNemar's test revealed significant differences (p = 0.016) when classifying IFX samples between therapeutic and supra-therapeutic ranges but no significant differences were found among the other ranges for either IFX or ADL. These results show that we should be cautious when using these rapid measurement methods, and new targets should probably be defined for IFX when using this new analytical method.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据