4.1 Article

Evaluating preanalytical sample storage parameters for nucleic acid-based detection of Neisseria gonorrhoeae

期刊

SEXUALLY TRANSMITTED INFECTIONS
卷 -, 期 -, 页码 -

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/sextrans-2023-055859

关键词

NEISSERIA GONORRHOEAE; GONORRHOEA; PUBLIC HEALTH; MICROBIOLOGY; Diagnostics

向作者/读者索取更多资源

This study assessed the impact of storage time, temperature, and buffer on the analytical sensitivity of detecting Neisseria gonorrhoeae. The results showed that N. gonorrhoeae could be successfully detected within a specific time range and temperature conditions.
Objectives To ensure accurate diagnosis of infectious diseases, preanalytical factors should be considered when assessing specimen quality and subsequent test result. Accordingly, we aimed to systematically assess the effect of storage time, temperature and buffer on the analytical sensitivity of detecting the sexually transmitted pathogen, Neisseria gonorrhoeae across multiple molecular diagnostic platforms.Methods Cultured N. gonorrhoeae was spiked into generic and commercial storage buffers and stored at four temperatures and five time points, ranging from -20 degrees C to 37 degrees C, over 30 days. Samples were processed using the Alinity m STI, Xpert CT/NG and Aptima Combo 2 nucleic acid amplification assays and an in-house quantitative PCR assay. A reduction in analytical sensitivity was defined as a significant (p<0.05) increase in cycle threshold (Ct) value relative to control samples.Results In total, 2756 samples were processed, with N. gonorrhoeae detected in 99.2% of samples. With respect to time, analytical sensitivity was maintained from day 2 (113/120; 94.2%) up to day 30 (110/120; 91.7%) relative to baseline samples. With respect to temperature, analytical sensitivity was maintained from -20 degrees C (147/150; 98.0%) up to 37 degrees C (136/150; 90.7%) relative to baseline samples. Generic buffers, Viral Transport Medium and Amies Liquid Media showed a reduction in analytical sensitivity compared with their commercial counterparts, Aptima Multitest Swab Transport Media and Abbott Alinity transport buffer using select diagnostic assays; this reduction appeared temperature dependent, with the largest differences in median Ct values observed at 37 degrees C (p<0.05).Conclusions Increased prevalence of sample self-collection for sexually transmitted infections (STIs) warrants an evaluation of preanalytical sample storage variables on diagnostic testing performance. Here, across a range of time points, temperatures and storage buffers, N. gonorrhoeae was successfully detected, supporting flexibility in sample storage, and by extension the feasibility of analysing self-collected samples to improve access to STI testing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据