A novel technique for measuring position resolution of Gas Electron Multipliers (GEM)

出版社

ELSEVIER
DOI: 10.1016/j.nima.2023.168836

关键词

Micro pattern gaseous detectors; Gas electron multipliers; Charge spread; Position resolution; Energy resolution; Gain uniformity

向作者/读者索取更多资源

The present study assesses the position resolution of Gas Electron Multiplier (GEM) detectors using a novel method. The results demonstrate that the double GEM can accurately determine positions in both x and y directions. Other standard characterizations were also evaluated and found consistent with existing literature.
Gas Electron Multiplier (GEM) detectors are widely used for tracking and imaging applications due to their excellent position resolution and high efficiency, among other factors. The present study aims to assess the position resolution by employing a novel method that utilizes a high-precision instrument for source movement. The results demonstrate that the double GEM can accurately determine positions with sigma values of up to 36.8 mu m and 54.6 mu m in the x and y directions, respectively. Additionally, other standard characterizations, such as energy resolution, gain uniformity, and charge spread, were evaluated experimentally and found to be consistent with existing literature. A Garfield simulation was also performed to validate the experimental charge spread results.