4.4 Article

Partially depleted operation of 250 μm-thick silicon carbide neutron detectors

出版社

ELSEVIER
DOI: 10.1016/j.nima.2023.168918

关键词

SiC detectors; Solid state detectors; Neutronics; Fusion plasma; Fusion diagnostics

向作者/读者索取更多资源

This paper presents the characterization of a series of SiC detectors operated with partial polarization, achieving excellent energy resolution and detection characteristics for both alpha particles and neutrons. The results demonstrate the potential of SiC detectors for fast neutron detection in fusion experiments. The limitations and challenges for manufacturing thick SiCs in the future are also discussed.
Fusion experiments put many challenges on the neutron detection methods, since they require fast neutron counters and spectrometers with good energy resolution capable of operating in harsh environments under a large range of neutron flux intensities. Recently, SiC detectors were proposed for these roles due to their outstanding properties. Among those, the operation of SiC with partial polarization is of interest since it might allow for a fast neutron detector with efficiency tunable online, which would allow a good functionality under a large range of flux intensities. This paper describes the characterization of a series of SiC detectors operated with a fraction of the polarization needed to achieve full depletion. The characterization is performed through I/V measurements, alpha irradiation in vacuum and neutron irradiation. A functionality with excellent detection characteristics is demonstrated, reaching an energy resolution of 1.05 % for 5.5 MeV alphas and 2.05 % for 14 MeV neutrons. Some limitations on the maximum bias will be also discussed, pointing out some challenges for the manufacturing of thick SiCs in the future.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据