Functionally graded 3D printed plates for rib fracture fixation

期刊

CLINICAL BIOMECHANICS
卷 111, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.clinbiomech.2023.106151

关键词

Additive manufacturing; Biomechanics; Fracture fixation; Rib fracture

向作者/读者索取更多资源

The study compared novel additively manufactured rib implants with a clinically used Control implant, finding that the novel implants had better construct stiffness and rotational performance than the control group, with all implants failing via bone fracture in the end.
Background: Design freedom offered by additive manufacturing allows for the implementation of functional gradients - where mechanical stiffness is decreased along the length of the implant. It is unclear if such changes will influence failure mechanisms in the context of rib fracture repair. We hypothesized that our novel functionally graded rib implants would be less stiff than controls and decrease occurrence of secondary fracture at implant ends.Methods: Five novel additively manufactured rib implants were tested along with a clinically used Control implant. Fracture reconstructions were modeled with custom synthetic rib bones with a transverse B1 fracture. Ribs were compressed in a cyclic two-point bend test for 360,000 cycles followed by a ramp to failure test. Differences in cyclic stiffness, 3D interfragmentary motions, ramp-to-failure stiffness, maximum load, and work to failure were determined.Findings: The Control group had lower construct stiffness (0.76 +/-; 0.28 N/mm), compared to all novel implant designs (means: 1.35-1.61 N/mm, p < 0.05) and rotated significantly more about the bending axis (2.7(degrees) +/- 1.3(degrees)) than the additively manufactured groups (means between 1.2(degrees) - 1.6(degrees), p < 0.05). All constructs failed via bone fracture at the most posterior screw hole. Experimental implants were stiffer than Controls, and there were few significant differences between functional gradient groups.Interpretation: Additively manufactured, functionally graded designs have the potential to change the form and function of trauma implants. Here, the impact of functional gradients was limited because implants had small cross-sectional areas.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据